Woodpecker Smart Ray dental portable x-ray machine-New release
{ "__shgImageV2Elements": { "uuid": "s-6534f54c-34c3-47ca-b4ca-95cc01fc06e3" } } Primary technical parameters Warranty User manual 1. Power adapter input:Power input: 100-240V~ 50/60Hz 1.1A2. Internal power supply: DC 10.8V3. Power adapter output: DC 15.0V 3.2A4. Type of radiation: X-Ray5. Electric powerMaximum output electric power: 0.14kw (70kV, 2mA)Nominal electric power: 0.14kw (70kV, 2mA, 0.1s)6. Tube voltage: tube voltage output is fixed at 70kV, error ±10%7. Tube current: tube current output is fixed at 2mA, error ±20%8. Loading time: the exposure loading time adjustment range is 0.02s ~ 2s, multipleexposure time settings adjustable according to R’10 numerical system; with deviation ±5% or ±20ms (whichever is greater)9. X- Ray tubeX-Ray tube model: D-045Focal spot: 0.4 mmTarget angle: 12.5°Total filtration: 1.5mmAl/70 kVPermanent filtration: 1.2mmAl/70 kVAdditional filtration: 0.3mmAl/70 kV10. Distance from focal spot