AimPlex™ EZPrep Filter Plate Washer
Supplier: AimPlex Biosciences, Inc. APPLICATION: The EZPrep Filter Plate Washer is designed to perform wash steps of AimPlex Bead-based Immunoassays while using the 96-well filter plate included in the Basic Kit. COMPONENTS: Washer, Tubing, ON/OFF Valve, 96-well Filter Plate. STORAGE: Room Temperature INSTALLATION: 1. Unpack the components and connect the supplied tubing to a 1- or 2-liter vacuum flask (not supplied) with the ON/OFF valve in between the Washer and vacuum flask. 2. For proper vacuum pressure adjustment, place the filter plate filled with 100 μL of ddH2O in each well on top of the Washer. 3. Open the On/Off valve to the ON position. 4. Press evenly on all four corners of the filter plate to form a tight seal. 5. Turn the vacuum ON then turn the vacuum OFF as soon as the liquid in the wells starts to go down. It should take 3-5 seconds to empty 100 μL of ddH2O from the wells with the remaining vacuum in the system after it is turned OFF. Vacuum pressure should be adjusted