Beckman Coulter Biomek NXp Automated Liquid Handling Workstation A31841
Beckman Coulter Biomek NXp Automated Liquid Handling Workstation Successful Software Communication & Plate Movement After connecting the NXp to the Biomek software, we performed the following tests: - NXp initializes and communicates with Biomek software: PASSED - Home all of the motors: PASSED - Prime syringe pumps: PASSED - Aspirate/dispense fluids: PASSED - Load/unload tips: PASSED - Lift/move microplates: PASSED - Create/edit/remove programs: PASSED You Will Receive: - Beckman Coulter Biomek NXp Automated Liquid Handler - MultiChannel 96 P20/200 Head with Gripper - PC with software installed - Monitor - Mouse, keyboard - Tip loader - Heating block with controller - Accuframe sensor - Accuframe tool - Metal reservoir with adapter plate - Serial cable - Power cables Consumables Not Included Consumables are required for this system to be used in production. You will need to acquire the consumables separately. They can generally be purchased from the manufacturer or a third-party dealer. Specifications Manufacturer Beckman Coulter Model Biomek NXp Catalog No. A31841 Serial No. A318410148 Environment Indoor use only Compressed Air Requirements NX-MC: Requires compressed airNX-S8: Only Span-8 instruments that use a Microplate Shaking ALP require compressed air Air Pressure: 60 - 100 psi of compressed air414 - 690 kPa of compressed airAir Flow: 2.25 cfm @ 60 psi 0.064 m3/min @ 414 kPa System Fluid Requirements NX-S8: Requires system fluid De-ionized or Distilled waterSystem fluid ambient operating temperature is 59°–86°F (15°–30°C).System fluid should be degassed for 24 hours prior to use. Weight 190 lbs. (86.2 kg) Ambient Operating Temperature 41°-86°F (5°-30°C) Humidity Restrictions <85% (non-condensing) @ 86°F (30°C) Altitude Restrictions up to 6562 ft (2000 m) Installation Category Category II Pollution Degree 2 Sound Pressure Level Maximum sound pressure: 82 dBMaximum sound pressure at 1 meter away: 75 dB Fuses US: 250 VAC, 3.15 amp, 5x20 mm, Slow-blow, UL recognized, CSA certifiedEurope: 250 VAC, 3.15 amp, 5x20 mm, Slow-blow, CENELEC approved Communications to Host RS-232 port Communications to Tip LoaderNX-MC: Uses a Tip Loader CAN Communications to Active ALPs via the Biomex NXp Instrument CAN Power Requirements 100-240V, 4A, 50/60Hz Overall Dimensions (WxDxH) 35 x 31 x 42"