LP870 Nanosecond Light Source
Testing optical receivers is harder than it looks. Getting accurate results is a lot easier if you're using a very fast, highly repeatable test source, because then the oscilloscope trace you're looking at is your receiver's performance, full stop. Our LP870 Nanosecond Light Source is a convenient and accurate low-coherence source for testing and characterization of silicon photodetection systems. Its fast edges (sub-4 ns rise and fall time typical) and clean pulses enable quick and reliable measurements up to 50 MHz bandwidth, with no guesswork. It provides a clean optical step input to measure the rise/fall times and overshoot of your system, which tells you its bandwidth as well. It's suitable both for lab use and for integration into test and calibration jigs working at up to 50 MHz. Using an incoherent source ensures eye safety and makes it a breeze to get uniform illumination of the test device. Just plug in the power cord and connect your pulse generator. The LP870 has high